Dr Pramod Kumar Pandey BSc (Hons), MSc, PhD, founder of PharmaGuru.co, is a highly experienced Analytical Research Expert with over 31 years in the pharmaceutical industry. He has played a key role in advancing innovation across leading Indian and global pharmaceutical companies. He can be reached at admin@pharmaguru.co

Isocratic elution and Gradient elution are two fundamental modes of mobile phase composition control in HPLC. In HPLC, isocratic elution maintains a constant mobile phase composition throughout the separation, while gradient elution involves progressively changing the mobile phase composition during the run. Gradient elution is typically preferred for complex mixtures with components that have a […]

Isocratic vs Gradient Elution In HPLC: How to Choose In 9 Minutes

Isocratic elution and Gradient elution are two fundamental modes of mobile phase composition control in HPLC. In HPLC, isocratic elution maintains a constant mobile phase composition throughout the separation, while gradient elution involves progressively changing the mobile phase composition during the run. Gradient elution is typically preferred for complex mixtures with components that have a wide range of polarities or retention times, as it provides better resolution and faster analysis. On the other hand, isocratic elution is simpler and is often used for separating mixtures with similar properties, such as compounds with similar polarities.

Isocratic vs Gradient Elution In HPLC: How to Choose
Isocratic vs Gradient Elution In HPLC (Image: CANVA)

Major Take away: FAQs

What is the difference between gradient and isocratic elution in HPLC?

In HPLC, isocratic elution uses a constant mobile phase composition throughout the separation, while gradient elution gradually changes the mobile phase composition during the run. Gradient elution is ideal for complex mixtures with varying polarities, offering better resolution and faster analysis, whereas isocratic elution is simpler and suited for mixtures with similar properties.

Why is gradient elution used in HPLC?

Gradient elution is used in HPLC to improve the separation of complex mixtures with components that have a wide range of polarities or retention times. It enhances resolution, speeds up analysis, and allows better separation of compounds that would be difficult to separate using isocratic elution.


Which elution mode will be suitable to separate two peaks?

Isocratic

Which elution mode will be suitable to separate 21 peaks?

Gradient

Which elution mode will be suitable to separate multiple co-eluting peaks?

Graientd

Isocratic vs Gradient Elution In HPLC

FeatureIsocratic ElutionGradient Elution
DefinitionUses a single, constant mobile phase composition throughout the run.Mobile phase composition is varied over time during the run.
Mobile Phase CompositionConstant throughout the analysis.Changes during the analysis (e.g., increase in solvent strength).
ComplexitySimpler and easier to set up.More complex and requires careful method development.
Separation EfficiencySuitable for separating compounds with similar polarities.Can separate compounds with a wide range of polarities.
Run TimeGenerally longer, as compounds with different affinities elute at different times.Can be shorter, as the gradient speeds up the elution of more strongly retained compounds.
Sample TypesBest for simple mixtures or when compounds are well-resolved with a single mobile phase.Ideal for complex mixtures, such as when analytes have a wide range of polarities.
ResolutionMay have lower resolution for complex mixtures.Higher resolution for complex mixtures with varied polarity.
Cost and EquipmentLess expensive due to simpler setup and equipment requirements.More expensive due to the need for sophisticated pumps and control systems.
ReproducibilityEasier to reproduce since the conditions are constant.Slightly harder to reproduce exactly due to the varying gradient.
ApplicationsSuitable for routine analyses, quality control, and simple samples.Used in complex analyses, such as in pharmaceutical and environmental testing.

1. Isocratic Elution

  • Definition: In isocratic elution, the composition of the mobile phase (the solvent or mixture of solvents) remains constant throughout the entire chromatographic run.
  • When to use:
    • Simpler samples: Best suited for samples with few components or when the analytes are well-separated using a single solvent or solvent mixture.
    • Less complex separation: Works well when there is little variation in the chemical properties (e.g., polarity) of the compounds you’re trying to separate.
    • Shorter run times: Since there is no change in mobile phase composition, you get faster and more predictable retention times.
    • More reproducible: As the mobile phase does not change during the run, you get highly reproducible separations if the system is stable.
  • Advantages:
    • Easier method development (since you don’t have to vary gradients).
    • Simpler to optimise the system for reproducibility.
    • Lower cost (since you only need one solvent or solvent mixture).
  • Disadvantages:
    • Limited flexibility if compounds in the sample have very different polarities.
    • May result in broad or co-eluting peaks if the analytes are too similar in polarity or interaction with the stationary phase.

2. Gradient Elution

  • Definition: In gradient elution, the mobile phase composition gradually changes during the chromatographic run. Typically, this means increasing the concentration of an organic solvent in a mixture (e.g., increasing acetonitrile or methanol in a water-based mobile phase).
  • When to use:
    • Complex samples: Useful for separating complex mixtures with a wide range of compounds of varying polarities.
    • Improving resolution: Gradient elution allows better separation for compounds that have a wide range of retention times, allowing late-eluting compounds to come off the column more quickly by increasing the solvent strength.
    • Better separation for poorly retained compounds: Compounds that would otherwise have very long retention times under isocratic conditions can be eluted faster in a gradient.
  • Advantages:
    • Greater versatility in separating complex mixtures.
    • Can handle a broad range of analyte polarities.
    • Better resolution for samples with components of very different retention times.
  • Disadvantages:
    • More complicated method development.
    • Potential for less reproducibility due to variations in gradient programming.
    • More expensive, as it may require multiple solvents.

You May Like

  1. Relative Response Factor (RRF) in Pharmaceutical Analysis
  2. How To Control Impurities In Pharmaceuticals: Get Mastery In 11 Minutes
  3. How To Calculate Potency, Purity and Assay In Pharmaceuticals

How to Choose?

  1. Sample Complexity:
    • Simple sample (few compounds): Isocratic elution is usually sufficient.
    • Complex sample (many compounds with different polarities): Gradient elution offers better separation.
  2. Compound Polarity:
    • Similar polarity: Use isocratic elution for ease and efficiency.
    • Wide range of polarities: Gradient elution is better to ensure that each compound has an optimal separation.
  3. Retention Time:
    • If analytes elute at very different times (some very early, some very late), a gradient approach can help achieve a more uniform run time.
    • If compounds have similar retention times and are well-separated, isocratic is the simpler choice.
  4. Speed of Analysis:
    • Isocratic elution is generally faster because the mobile phase composition does not change.
    • Gradient elution might take longer, as the solvent composition is adjusted during the run.
  5. Resolution Requirements:
    • If you need to achieve high resolution for a complex sample, gradient elution is often the better choice. It allows better optimisation of separation.
  6. Reproducibility:
    • Isocratic elution is typically more reproducible since the conditions stay constant throughout the run.
    • Gradient elution can introduce variability, especially if the gradient is not properly controlled.

Isocratic elution and Gradient examples

  • Isocratic:
    • Water-soluble small molecules with similar hydrophobicity or size (e.g., for routine quality control of a single compound in pharmaceutical analysis).
  • Gradient:
    • Peptide or protein analysis (where a wide range of polarities and sizes exist).
    • Complex environmental samples (such as soil or water extracts containing various organic compounds with different polarities).

Case Study: Selecting Between Isocratic and Gradient Elution in HPLC

Background: A pharmaceutical company is developing a new drug, a mixture of several compounds, with varying polarities, including the active pharmaceutical ingredient (API) and some impurities that need to be quantified during manufacturing. The company needs to choose an appropriate HPLC method to evaluate the purity and quality of the drug. The drug formulation contains the API, excipients, and several degradation products formed during stability testing.

The goal is to identify and quantify the API and impurities with high resolution, using the most cost-effective and reproducible method.

Step 1: Understanding the Sample Composition

  • Active Pharmaceutical Ingredient (API): Moderately hydrophobic.
  • Impurities: Varying degrees of polarity, ranging from very hydrophilic (polar) to more lipophilic (non-polar).
  • Excipients: Generally non-polar, but do not interfere with the analytes’ retention during HPLC.

The compounds have varying molecular weights and polarities, and some degradation products are expected to have retention times that are much longer than others.

Step 2: Initial Evaluation of Elution Mode

  • Isocratic Elution:
    • Since the API and some degradation products have moderately similar polarities, an isocratic method could potentially work. However, for impurities that differ greatly in polarity, some compounds might either co-elute or require longer retention times, leading to poor separation.
    • Isocratic might not achieve sufficient separation between the API and the impurities (especially the highly polar degradation products).
    • The lack of flexibility for handling a range of retention times could compromise the analysis.
  • Gradient Elution:
    • Gradient elution would allow the separation of the API from both the polar impurities and the non-polar degradation products by gradually increasing the strength of the mobile phase (e.g., from 10% acetonitrile to 50% over 15 minutes).
    • This would shorten retention times for the highly retained compounds (e.g., degradation products) while allowing the polar impurities to elute early with a less aggressive solvent composition.
    • The gradient would ensure that all compounds in the sample, with their varying polarities and retention times, are optimally separated.

Step 3: Testing the Elution Methods

  • Isocratic Test:
    • Method: Mobile phase = 50% water + 50% acetonitrile.
    • Observation: The API elutes well, but some impurities show poor resolution, especially a polar degradation product that co-elutes with another impurity.
    • The retention time for the API is acceptable, but the impurities show broad peaks that make quantification difficult.
  • Gradient Test:
    • Method: Start with 10% acetonitrile (water-based) and increase to 50% acetonitrile over 15 minutes.
    • Observation: The polar impurities elute early, while the API and non-polar degradation products elute later. The separation is improved, and all peaks are well-defined with sharper, more resolved peaks.
    • The non-polar impurities now elute more efficiently, with the degradation products being separated based on their retention times.

Step 4: Decision on Elution Mode

After conducting the tests, the gradient elution method is deemed the most suitable choice for the following reasons:

  1. Improved Separation: The gradient allows optimal separation of the API, impurities, and degradation products by adjusting the mobile phase to different solvent strengths over time. This results in sharper, more defined peaks for each analyte.
  2. Resolution of Polar Impurities: The gradient helps separate even highly polar degradation products, which would otherwise co-elute in an isocratic system.
  3. Efficient Run Time: Despite the extended run time compared to an isocratic method, the gradient ensures a faster and more efficient separation for all components in the sample.
  4. Reproducibility: The gradient profile is stable, allowing for reproducible results across different runs, ensuring the integrity of the analysis.

Step 5: Final Method and Parameters

  • Mobile Phase:
    • Solvent A: Water (with 0.1% trifluoroacetic acid).
    • Solvent B: Acetonitrile (with 0.1% trifluoroacetic acid).
  • Gradient Profile:
    • 0-5 min: 10% B → 30% B
    • 5-10 min: 30% B → 50% B
    • 10-15 min: 50% B → 70% B
    • 15-20 min: 70% B → 100% B
    • 20-25 min: 100% B → 10% B (re-equilibration)
  • Flow Rate: 1.0 mL/min
  • Detector: UV at 254 nm
  • Column: C18, 150 mm x 4.6 mm, 5 µm particle size.

Step 6: Validation and Outcome

  • The gradient method is validated for accuracy, precision, and robustness.
  • Linearity and range are established, with good separation of the API and all degradation products, ensuring compliance with pharmaceutical standards.
  • Quantitative analysis is reliable, and the results are reproducible across different batches of the drug.

Note: For this pharmaceutical case, gradient elution was the optimal choice due to the complex nature of the sample, the need for high resolution, and the varied polarities of the compounds.

Conclusion

  • Choose isocratic elution when you have simple samples or if the analytes have similar properties. It’s quicker and easier to set up.
  • Opt for gradient elution for complex samples with a wide range of analyte polarities. This method offers better separation and resolution but at the cost of increased method complexity and longer analysis time.

Further Reading

  1. LIQUID CHROMATOGRAPHY– MASS SPECTROMETRY: Robert E. Ardrey
  2. HPLC METHODS FORRECENTLY APPROVED PHARMACEUTICALS: George
  3. <621> Chromatography
About Dr Pramod Kumar Pandey
Dr Pramod Kumar Pandey

Dr Pramod Kumar Pandey BSc (Hons), MSc, PhD, founder of PharmaGuru.co, is a highly experienced Analytical Research Expert with over 31 years in the pharmaceutical industry. He has played a key role in advancing innovation across leading Indian and global pharmaceutical companies. He can be reached at admin@pharmaguru.co