Learn the importance of solution stability in analytical method validation with a practical case study. Understand why and how this test is performed, and gain the skills to apply it independently
Understanding Solution Stability in Analytical Method Validation: A Practical Case Study
Solution Stability In Analytical Method Validation is a critical parameter, ensuring the reliability and accuracy of analytical results over time. In this article, I will explore the importance of solution stability testing, explain the methodology used to assess it, and present a real-world case study to illustrate its practical application. By the end of this post, you will not only understand the purpose of solution stability testing but also gain the confidence to perform it independently and effectively in your own laboratory work.
Solution stability is a crucial component of analytical method validation, ensuring that analytical results remain accurate and reliable over time. It involves evaluating how long standard and sample solutions can be stored under specific conditions – such as room temperature or refrigeration – without compromising their integrity. This is typically assessed by comparing the results of freshly prepared solutions with those stored over defined time intervals, allowing analysts to determine the acceptable duration for solution use in routine analysis..
During HPLC or GC analysis, several samples are prepared and analysed sequentially. Analysis may take several days to complete depending upon the run time and number of samples. All analysis result will be invalid if the standard or samples are not stable. To avoid such failure, stability of solution is demonstrated.
You may like:
For assay test
The standard/sample solution is tested for up to 5 days by comparison of the response factor of an old solution (injected as a sample) to freshly prepared standard solutions.
The test for standard solution stability is also valid for sample solution stability because same diluent is used for sample-solution preparations
Acceptance criteria: The test passes the acceptance criteria if % RF (Response factor) difference between freshly prepared standard solution and stability solution is ≤ 2%.
For Related substances test
The sample solution is tested for up to 5 days for related substances stability testing. Result of old samples solution (stability solution) is compared with the freshly prepared sample solutions.
Acceptance criteria: No new peak higher or equal to QL (Quantitation limit) should appear in the old sample solutions
Use the following steps when testing solutions stability test for an assay:
Solution | V0 | V1 | V2 | V3 | V4 | V5 |
---|---|---|---|---|---|---|
Time interval (hours) | 0 (initial) | 12 | 24 | 36 | 48 | 60 |
A drug substance D having the following specifications for related substances and assay test:
Assay : Not less than 98.0%
Sample nominal concentration (assay analysis concentration) in the method is 1.0 mg/ml .
Let us demonstrate solution stability at different intervals:
Solution stability at 12 hours
Freshly prepared standard solution (V0)
Injection | Area response | RF (Area response/1.025) |
1 | 154570 | 150800 |
2 | 154560 | 150790 |
3 | 154050 | 150293 |
4 | 154065 | 150307 |
5 | 154072 | 159314 |
6 | 154075 | 150317 |
Average (A) | NA | 151970 |
V1 solution
Injection | Area response | RF (Area response/1.020) |
1 | 154573 | 151542 |
2 | 154566 | 151535 |
Average (V1) | NA | 151539 |
% RF difference between freshly prepared standard solution and stability solution V1 (after 12 hours)
% RF Difference = [(|151970 – 151539|)/151755 ]x 100 = 431/151755 × 100 = 0.28%
Injection | Average RF of V0 | Average RF of V1 | % RF difference between F and V1 solution |
Average (V1) | 151970 | 151539 | 0.28% |
Conclusion
The test passess the acceptance criteria at 12 hours since % RF difference between
V0 and V1 solution is 0.28% < 2%
Note:
Solution | V0 | V1 | V2 | V3 | V4 | V5 |
---|---|---|---|---|---|---|
Time interval (hours) | 0 (initial) | 12 | 24 | 36 | 48 | 60 |
A drug substance D having the following specifications for related substances and assay test:
Related substances test (by area normalisation method):
Sample nominal concentration (analysis concentration) in the method is 1.0 mg/ml. QL of the method is 0.04%.
Let us demonstrate solution stability at different intervals:
Solution stability at 12 hours
Freshly prepared standard solution
Impurities | Related substances at 0 hour | Related substances at 12 hours | % Difference |
Impurity A | 0.15% | 0.16% | 6.5% |
Any unknown impurity | <0.04% | <0.04% | NA |
Total impurity | 0.15% | 0.16% | 6.5% |
Conclusion
The test passess the acceptance criteria at 12 hours since the % difference of impurity A is 6.5% (< 10%) and not any new impurity (having value ≤ QL) forms in the chromatogram.
Note: Similarly perform solution stability for V2, V3, V4, V5 solution and conclude the result
Solution stability data is very helpful in routine analyses especially when analysis is performed for longer time in sequence. It prevents failure of result solution stability analytical error.
I hope this article has helped you understand solution stability and its importance in pharmaceutical analysis. Now you can independently perform solution testing during method development and method validation.
You may also want to check out other articles on my blog, such as:
If you found this article helpful, then please subscribe to our Newsletter and also write your comment. You can also follow me on LinkedIn, YouTube, Facebook and Twitter.
During Chromatographic analysis, several samples are prepared and analysed sequentially. Analysis may take several days to complete depending upon the run time and number of samples. All analysis result will be invalid if the standard or samples are not stable. To avoid such failure, stability of solution is demonstrated
Samples and solutions are prepared as per method given in the monograph. Solution are injected at different time interval such as 0 hour, 12 hour, 24 hours, 36 hours, 48 hours and 60 hours. The % differance of result is calculated at different time interval between freshly prepared solution and stability solution. The time period for which result remains within the acceptance criteria is called solutions stability.
For assay, the % response factor difference between freshly prepared solution and stability solution ( after a particular given period) must be less than 2.0%.
References
Abbreviations
Disclaimer
The numerical data used in the tables or calculations are not actual data. It is designed to explain the topic.
Subscribe to Stay Informed with the Latest Industry Insights!
Quick Links