Follow Us :

Pharmaceutical Solutions through Blogs and Courses

Blog

By Dr. Pramod Kr. Pandey - June 9, 2025

Dr. Pramod Kr. Pandey is a distinguished Analytical Research Expert with over three decades of extensive experience in the pharmaceutical industry. He has contributed his expertise to both leading Indian and global pharmaceutical companies, consistently driving innovation and excellence in analytical research

Learn HPLC method development from a chromatography expert with 30 years of experience. Includes practical tips, case studies, and FAQs

Fast and Cost Effetive HPLC Method Development In 18 Easy Steps

“A good HPLC method doesn’t happen by chance, it’s designed with purpose, patience, and practice.”

Method development in HPLC means creating a plan to accurately and efficiently separate and analyse compounds, while ensuring the method is reliable, practical, affordable, and scientifically sound. It is more than just adjusting parameters, it’s a scientific art that blends analytical chemistry, organic chemistry, and real-world laboratory experience

That’s exactly why I’ve decided to share insights gained over 30 years of practical experience, distilled into just 15 minutes of reading.

In this article, you’ll discover a step-by-step guide to HPLC method development, covering everything from initial planning and separation optimisation to detection, finalisation, and real-world case studies. I’ve also included a handy FAQ section to address common questions that arise during method development.

Whether you’re a beginner or a working professional refining your approach, this guide is designed to help you navigate the process more efficiently and with greater confidence

HPLC Method Development

HPLC method development involves creating a reliable, practical, and validated analytical approach tailored to the chemical properties of the analyte, while considering cost and operational feasibility

Related: How To Control Impurities In Pharmaceuticals: Get Mastery In …

HPLC Method Development Steps

The following is a 16-step process for HPLC method development:

  1. To Know about the Goal of the Method
  2. To Know Details about the sample
  3. Collection of Literature or Literature Report
  4. HPLC Mode Selection
  5. Column Selection
  6. Mobile Phase Selection
  7. Detector Selection
  8. Elution Mode Selection
  9. Method Optimisation
  10. SST (System Suitability Test)
  11. Sample preparation procedure
  12. Flow Rate Optimisation
  13. Column Temperature Optimisation
  14. Mode of Calculation
  15. Adjust and Refine Chromatographic Conditions
  16. Method Verification or Method Validation
  17. Documentation and Reporting
  18. Method Transfer

1. To Know about the Goal of the Method

In this step, find out the goal of the method, such as whether you are going to develop a method for identity testing, purity testing, assay testing, related substance testing, chiral purity or limit testing. What will be the specification? What will be the method of calculation?

Types of test

  • Qualitative test
    • Identification test
    • Monitoring the reaction mixture
    • Purity test
    • Impurity profile test or Related substances test
  • Quantitative test
    • Assay test
    • Impurity profile test or Related substances test
    • Content test

2. To Know Details about the sample

Get the following sample details in this step:

  1. Route of Synthesis (ROS): Collect ROS of all purchasing materials like starting materials, key starting materials, reagents, chemicals and final API (Active Pharmaceutical Ingredients).
  2. Possible impurities and degradants: Collect information about possible impurities and degradation products with the help of ROS and literature reports.
    • Note: Find out all the concerned impurities related to a method with the help of literature and write it down in tabular form. It is required to prove the selectivity and specificity of the HPLC method which you are going to develop.
  3. Literature reported impurities: Collect all the published data and make the literature report. This report contains a piece of information related to impurities, chromatographic conditions, pH, pKa, stability, etc.
  4. Stability of the sample and impurities (under certain pH, temperature, or light): Collect all the information related to the stability of the main analyte and impurities using ROS and literature report. It is very helpful during diluent selection
  5. Structures of the main analyte and impurities: n this step, write down the structure, molecular weight and polarity nature of the molecule in tabular form. It is very helpful during column selection and mobile phase selection during method development.
  6. Standard availability: Collect all the parameters of related impurities and main analysis as without this method, development is impossible. If the standard is not available, then a marker or contaminated sample (containing relevant impurities) can be used.
  7. Solubility of the sample and impurities: Collect information related to the solubility of the sample and impurities with the help of chemical R&D scientists and literature. For any new compound where solubility information is not available, conduct a solubility study. This is very helpful in Diluent selection during sample preparation
  8. Understand chemical properties: polarity (Hydrophilicity/hydrophobicity), pKa, solubility, UV absorbance: Collect information related to Hydrophilicity and hydrophobicity for concerned impurities and main analyte with the help of literature. It is very helpful in the pH selection of the mobile phase and column selection. Also, collect information related to pK value for the concerned impurities and the main analyte with the help of literature. If not available, then find out the same using titration or UV method. It is very helpful in pH selection of the mobile phase.

3. Collection of Literature or Literature Report

Like other Analytical methods, HPLC method development typically begins with a comprehensive literature survey. This involves reviewing various pharmacopoeias such as the Indian Pharmacopoeia (IP), United States Pharmacopoeia (USP), European Pharmacopoeia (EP), British Pharmacopoeia (BP), and Japanese Pharmacopoeia (JP), as well as consulting relevant chromatographic and scientific journals. The goal is to identify any existing analytical methods that may be suitable for the intended application. Even if a suitable method is identified, it is essential to carry out method development, optimisation, and validation to ensure the method is robust, reproducible, and fit for its intended purpose. This process may involve adapting and improving the existing method to meet specific analytical requirements and regulatory standards

4. HPLC Mode Selection

In this step, select the appropriate HPLC mode – such as reversed-phase chromatography (RPC), normal-phase chromatography (NPC), or another specialised mode – based on the polarity, solubility, and chemical nature of the analyte and its impurities. Reversed-phase chromatography is generally the preferred starting point due to its broad applicability, ease of use, and compatibility with a wide range of compounds. In most cases, I recommend beginning with RPC unless the analyte properties strongly suggest otherwise.

Select the most appropriate chromatography mode based on the sample and its impurity properties:

  • Normal Phase HPLC: For polar analytes.
  • Reverse Phase HPLC (RP-HPLC): For non-polar or moderately polar analytes.
  • Ion-Exchange HPLC: For charged analytes (typically used for peptides, proteins, or inorganic ions).
  • Size-Exclusion HPLC: For large molecules like proteins or polymers.
  • Affinity Chromatography: For specific binding interactions (e.g., antibodies or specific ligands).

5. Column Selection

It doesn’t matter how sophisticated and expensive your HPLC system is, you can’t perform method development if you haven’t chosen the right column. The general rule is to use a polar column for a polar molecule and a non-polar column for a non-polar molecule.

  • Column Chemistry: Select an appropriate column based on the mode chosen (e.g., C18, C8 for RP-HPLC, silica-based for normal phase, or ion-exchange columns for ionic compounds).
  • Particle Size: Smaller particle sizes (e.g., 1.8-5 µm) improve resolution but can increase backpressure.
  • Column Dimensions: The column length and inner diameter affect separation efficiency. Common sizes are 4.6 mm × 150 mm or 4.6 mm × 250 mm.

6. Mobile Phase Selection

Mobile Phase Selection plays a vital role in HPLC method development. The general rule is to use an acidic mobile phase for an acidic molecule, a basic mobile phase for a basic molecule and a neutral mobile phase for a non-polar molecule.

  • Solvent System: The mobile phase needs to match the polarity of the analytes. In RPC, a gradient or isocratic system using water and organic solvents like methanol or acetonitrile is common.
  • Buffer Selection: If working with ionic compounds or adjusting pH, use an appropriate buffer (e.g., phosphate, acetate) to maintain consistent pH.
  • pH Control: For stability and reproducibility, choose a mobile phase with a stable pH. For example, in RP-HPLC, the pH is usually maintained between 2-8.
CompoundsMobile Phase
NeutralCH3COONH4 buffer pH 9.0 and organic solvents such as Acetonitrile, Methanol, Ethanol, Isopropyl alcohol, etc
AcidicA mixture of H2O and organic solvents such as Acetonitrile, Methanol, Ethanol, Isopropyl alcohol, etc.
AcidicA mixture of H2O of 0.02 to 0.1M KH2PO4 or NaH2PO4 and organic solvents such as Acetonitrile, Methanol, Ethanol, Isopropyl alcohol etc
Basic10- 20 mM Na2HPO4 or K2HPO4 with pH 8.0 and organic solvents such as Acetonitrile, Methanol, Ethanol, Isopropyl alcohol, etc
BasicA mixture of H2O of 0.02 to 0.1M KH2PO4 or NaH2PO4 and organic solvents such as Acetonitrile, Methanol, Ethanol, Isopropyl alcohol, etc
Highly acidicUse basic ion pairs such as tetrabutylammonium hydroxide (TBA) in the mobile phase
Highly basicUse acidic ion pairs, such as alkyl sulfonate sodium salt, in the mobile phase

7. Detector Selection

The commonly used detectors in pharmaceutical industries are:

  • Ultra-violet/UV detecto
  • Mass spectrometer(MS) detector and
  • Refractive Index (RI) detector

Among the above detectors, the UV detector is widely used in industries. Mass detectors are extremely expensive and are used for structure elucidation and quantification at very low levels.The RI detector is used for pharmaceuticals which do not have any reaction, like glucose, sucrose, etc.

Note: Pick based on analyte characteristics and required sensitivity

8. Elution Mode Selection

The following elution modes are widely used in HPLC analysis:

  • Isocratic mode and
  • Gradient Mode

If your sample has components of almost similar polarity and the number of components is 3 to 5, then you can use the isocratic mode. If your sample has multiple components of different polarity, you can use gradient mode.

9. Method Optimisation

  • Retention Time: Adjust the mobile phase composition, flow rate, or column temperature to fine-tune the retention times and ensure good separation of analytes.
  • Peak Resolution: Aim for a resolution (Rs) of ≥1.5 for baseline separation. Modify parameters such as the gradient program, mobile phase pH, or ion-pairing agents to improve peak separation.
  • Sensitivity: Adjust the detection wavelength or method parameters (e.g., UV wavelength, flow rate) to improve sensitivity if needed.

10. SST (System Suitability Test)

The selection of SST acceptance criteria plays an important role in HPLC method development as it tells about the performance of the HPLC column and the HPL system for the intended analyte. Wrong SST selection may result in failure of the pharmaceuticals. The commonly used SST parameters are:

  • Precision: Precision is mostly used for assay tests, content tests and related substances tests. RSD of 6 injections should be kept in the SST, and it should be based on the specification
  • QL (Quantification Limit): QL is part of SST for Related substances test/Impurity profile test/content test (at low level), but not assay test. SST evaluation criteria S/N≥ 10
    • Note: Some pharmacopoeias use a sensitivity test in place of the QL test. Both QL and sensitivity tests are the same
  • Resolution (R): When peaks are eluting close to each other, then R is kept as one of the SST acceptance criteria. The General rule for keeping the R in SST:
    • R must be kept between two close eluting peaks
    • R can be kept between the analyte peak and the impurity peak
    • R can be kept between the Impurity peak and the Analyte peak
    • R can be kept between the Impurity peak and the Impurity peak
    • R≥ 2 (with baseline separation between the adjacent peaks
    • R can also be kept less than 2 with scientific justification
  • Theoretical plate or Column Efficiency (N): The following are the conditions for keeping the N in SST:
    • When peaks are eluting far away, then N is kept as one of the SST acceptance criteria
    • The limit must be decided based on trend data
    • N≥ 5000 for HPLC and N ≥ 10000 for GC
    • A lower N limit can be considered with scientific justification
    • Generally, N of the main peak is kept as an SST acceptance criterion
  • Tailing Factor(T): The following are the conditions for keeping the N in SST:
    • When peaks are eluting far away, then N is kept as one of the SST acceptance criteria
    • The limit must be decided based on trend data
    • N≥ 5000 for HPLC and N ≥ 10000 for GC
    • A lower N limit can be considered with scientific justification
    • Generally, N of the main peak is kept as an SST acceptance criterion

11. Sample preparation procedure

Sample preparation involves the following steps:

  • Selection of solvent to use as a diluent: Select the solvent in which the sample and its impurities are soluble. Organic solvents such as acetonitrile, methanol, ethanol, isopropyl alcohol or their mixtures or their mixtures with aqueous can be used.
  • Compatibility of diluent with mobile phase and wavelength: Diluent must be compatible with the mobile phase and should not be reactive with the column or sample components
  • Concentration: The sample concentration should be adjusted so that the column does not become oversaturated. Also, ensure that the analyte concentration falls within the linear dynamic range of the detector.
  • Stirring or Sonication: Stirring or sonication time and temperature should be properly evaluated to avoid any degradation of impurities or the sample
  • Sample Clean-up: If the sample matrix is complex, it may require filtration, solid-phase extraction (SPE), or protein precipitation to minimise matrix interference.
  • Injection Volume: A typical volume for HPLC injections is between 5 µL and 20 µL, but this depends on the detector and column
  • Filtration: Filter the sample solution to remove undissolved matter. Filtration is generally performed for dosages from samples which contain excipients such as tablets, capsules, syrups etc. Filter effects must be evaluated during development.

12. Flow Rate Optimisation

  • Flow Rate: A typical starting flow rate for a 4.6 mm column is 1 mL/min. This can be adjusted depending on the system’s backpressure and the separation needs.
  • Column Efficiency: Higher flow rates reduce resolution, while lower flow rates might improve resolution but also increase analysis time.

13. Column Temperature Optimisation

Most of the HPLC columns are stable up to 800C. But try to optimise the HPLC column temperature between 10 to 350C °C because at higher temperatures the life of the column decreases.

14. Mode of Calculation

In this step mode of calculation is selected. In HPLC, the following modes of calculation are widely used:

  • Area% or Area normalisation method and
  • External standard Method

External standard is used for quantitative tests such as assay, related substances and content tests whereas area normalisation is used for qualitative analysis such as purity tests, reaction monitoring tests and impurity profile tests.

15. Adjust and Refine Chromatographic Conditions

  • Adjust gradient, pH, flow, or temperature to separate overlapping peaks.
  • Try column alternatives if the resolution is poor.

16. Method Verification or Method Validation

Perform method validation or mini-validation so that the method can be validated when needed and any surprises can be avoided. The verification parameters can be selected based on the sensitivity of the method, which may include:

  • Specificity: Ability to separate analytes from matrix components.
  • Linearity: The method’s ability to produce a straight-line response over the concentration range.
  • Accuracy: The closeness of measured values to the true value.
  • Precision: Repeatability and reproducibility of results.
  • Detection Limit (DL) and Quantification Limit (QL).
  • Robustness: The method’s reliability under small variations in conditions (e.g., slight changes in temperature or mobile phase composition).

17. Documentation and Reporting

  • Method Description: Document all parameters used, such as column type, mobile phase, flow rate, detection conditions, and sample preparation procedures.
  • Standard Operating Procedures (SOPs): Write SOPs to ensure the method is reproducible by different operators or in different laboratories.

18. Method Transfer

Method Transfer step includes:

  • Transfer of the method to the Quality control lab or end user lab
  • Routine monitoring and maintenance
  • Regularly monitor system performance to ensure stability.
  • Perform routine maintenance on the HPLC system (e.g., cleaning, replacing filters, and inspecting pumps).

Expert Tips on HPLC Method Development

  • Start with default conditions, then optimise step-by-step.
  • Keep the method simple unless complexity is required.
  • Document every trial—failures teach as much as successes.
  • Always check system suitability before moving to validation.

Case Study-1: Separation of Impurities in an API

How to develop a Related substance/ Impurity profile test method for N-(4-Hydroxyphenyl)acetamide (Paracetamol) containing the following ROS:

Method Development Steps

  1. Type of Test: Related Substances/Impurity profile test
  2. To know about the sample
    • ROS: Available
    • Possible impurities: Phenol, 4-Nitrophenol, 4-aminophenol
    • Maine analyte: N-(4-Hydroxyphenyl)acetamide
    • Main analyte and impurities: Both the main analyte and impurities are soluble in a mixture of water and acetonitrile
  3. HPLC Mode Selection: Reverse phase mode
  4. Column Selection: Both the main analyte and impurities are separated on the C18 and C8 columns. But C18 was selected for routine analysis.
  5. Mobile Phase Selection: All analytes are well-separated in acidic pH (pH 3 to 4) in the mobile phase that contains 0.02M KH2PO4 and acetonitrile
  6. Detector Selection: Since all analytes can give good absorbance therefore a UV detector is the best choice. The PDA detector suggests a 230 nm operating wavelength as all analytes have good absorption at this wavelength.
  7. Elution Mode Selection: Since compounds have different polarities and hence linear gradient mode was selected for this test.

*Typical gradient design (not actual)

Time (in minute)A( 0.02M KH2PO4)B (Acetonitrile)
07525
212575
272575
27.57525
347525
  1. Method Optimisation: The method has been optimised using adjustments to column temperature, flow rate, and sample concentration.
  2. SST (System Suitability Test) acceptance Criteria Selection: Amino phenol and phenol are the nearest eluting peaks, and hence, R is kept between these two as the first SST acceptance criteria. The Tailing factor has been kept as the second SST acceptance criterion
  3. Sample preparation procedure: Analytes found soluble in the mixture of water and acetonitrile (60:40)
  4. Mode of Calculation: The area normalisation method using relative response factor(RRF) has been selected as a mode of calculation
  5. Method Verification/Mini Validation: Mini validation confirms the suitability of this method

The following is the elution pattern of the final optimised related substances method:

  1. 4-aminophenol
  2. Phenol
  3. N-(4-Hydroxyphenyl)acetamide
  4. 4-Nitrophenol
HPLC Method Development

Case Study-1: Separation of Nonpolar Compounds:

Common HPLC Method Development Troubleshooting

Case Study 2: API Assay Development

  • Problem: API and impurity co-elution
  • Solution: Switched from methanol to acetonitrile; adjusted gradient
  • Outcome: Resolution improved from 1.2 to 2.8

Case Study 3: Forced Degradation Stability Testing

  • Challenge: Degradation peaks interfering with the analyte
  • Solution: Buffer pH adjusted from 4.5 to 3.0; used C8 column
  • Result: Achieved complete baseline separation

Conclusion

HPLC method development is a critical and innovative analytical skill that requires a systematic approach to fine-tuning multiple parameters to achieve optimal separation, sensitivity, and reproducibility for specific analytes. From defining the analytical objective to method validation and thorough documentation, every step plays a vital role in ensuring the method is scientifically sound and practically applicable.

I hope this article has provided you with a clear and structured understanding of the HPLC method development strategy, empowering you to approach your next project with greater confidence and precision.

Related:

FAQs

How are HPLC methods developed?

Using proper column selection, mobile phase selection and sample preparation procedure HPLC can be developed

How to do HPLC step by step?

HPLC step-by-step procedure involves HPLC mode selection, column selection, mobile phase selection, detector selection and elution mode selection, SST selection and sample preparation procedure

What are the principles of HPLC method development?

The process of creating a vision-based plan for developing an HPLC method, considering the cost, reliability, practicality, and validity of the method, is called method development strategy.

How do I choose between acetonitrile and methanol?

Try both. ACN often gives sharper peaks and lower backpressure.

What if my analyte has no UV absorbance?

Use derivatisation or switch to detectors like ELSD or MS

Can I use water without a buffer?

Not recommended. Buffers control pH and improve reproducibility.

How long should a method take to develop?

It depends—but expect 2 to 6 weeks for robust methods.

Further Reading:

  • Practical High-Performance Liquid Chromatography: VERONIKA R. MEYER
  • ICH Q14

Abbreviations

  • SST: System suitability test
  • HPLC: High-pressure liquid chromatography
  • R&D: Research and Development

Disclaimer

To explain the topic, some data has been designed

About Dr Pramod Kumar Pandey
Dr. Pramod Kr. Pandey

Dr. Pramod Kr. Pandey is a distinguished Analytical Research Expert with over three decades of extensive experience in the pharmaceutical industry. He has contributed his expertise to both leading Indian and global pharmaceutical companies, consistently driving innovation and excellence in analytical research

Subscribe to Stay Informed with the Latest Industry Insights!

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!
×